Islet Autoantibody Standardization Program: interlaboratory comparison of insulin autoantibody assay performance in 2018 and 2020 workshops

28 3月 2023
Author :  

Ilaria Marzinotto1 iD David L. Pittman2 iD Alistair J. K. Williams3 iD Anna E. Long3 iD Peter Achenbach4 iD Michael Schlosser5,6 & Beena Akolkar7 iD William E. Winter2 & Vito Lampasona1 iD participating laboratories

Received: 30 August 2022 /Accepted: 21 December 2022 / The Author(s) 2023

Abstract

Aims/hypothesis The Islet Autoantibody Standardization Program (IASP) aims to improve the performance of immunoassays measuring autoantibodies in type 1 diabetes and the concordance of results across laboratories. IASP organises international workshops distributing anonymised serum samples to participating laboratories and centralises the collection and analysis of results. In this report, we describe the results of assays measuring IAA submitted to the IASP 2018 and 2020 workshops. Methods The IASP distributed uniquely coded sera from individuals with new-onset type 1 diabetes, multiple islet autoantibodypositive individuals, and diabetes-free blood donors in both 2018 and 2020. Serial dilutions of the anti-insulin mouse monoclonal antibody HUI-018 were also included. Sensitivity, specificity, area under the receiver operating characteristic curve (ROCAUC), partial ROC-AUC at 95% specificity (pAUC95) and concordance of qualitative/quantitative results were compared across Results Results from 45 IAA assays of seven different formats and from 37 IAA assays of six different formats were submitted to the IASP in 2018 and 2020, respectively. The median ROC-AUC was 0.736 (IQR 0.617–0.803) and 0.790 (IQR 0.730–0.836), while the median pAUC95 was 0.016 (IQR 0.004–0.021) and 0.023 (IQR 0.014–0.026) in the 2018 and 2020 workshops, respectively. Assays largely differed in AUC (IASP 2018 range 0.232–0.874; IASP 2020 range 0.379–0.924) and pAUC95 (IASP 2018 and IASP 2020 range 0–0.032). Conclusions/interpretation Assay formats submitted to this study showed heterogeneous performance. Despite the high variability across laboratories, the in-house radiobinding assay (RBA) remains the gold standard for IAA measurement. However, novel non-radioactive IAA immunoassays showed a good performance and, if further improved, might be considered valid alternatives to RBAs.

Keywords Autoantibodies . IAA . IASP interlaboratory comparison study . Sensitivity . Specificity . Type 1 diabetes

Ilaria Marzinotto and David L. Pittman contributed equally.

Alistair John Kerr Williams died on 17th September 2020.

* Peter Achenbach

该Email地址已收到反垃圾邮件插件保护。要显示它您需要在浏览器中启用JavaScript。

* Vito Lampasona

该Email地址已收到反垃圾邮件插件保护。要显示它您需要在浏览器中启用JavaScript。

1 San Raffaele Diabetes Research Institute, San Raffaele Scientific Institute, Milan, Italy

2 Department of Pathology, University of Florida, Gainesville, FL, USA

3 Diabetes and Metabolism, Translational Health Sciences, University of Bristol, Bristol, UK

4 Institute of Diabetes Research, Helmholtz Munich, German Research Center for Environmental Health, Neuherberg, Germany

5 Department of General Surgery, Visceral, Thoracic and Vascular Surgery, University Medical Center Greifswald, Greifswald, Germany

6 Institute of Pathophysiology, Research Group of Predictive Diagnostics, University Medical Center Greifswald, Karlsburg, Germany

7 Division of Diabetes, Endocrinology, and Metabolic Diseases, National Institute of Diabetes and Digestive and Kidney Diseases, Bethesda, MD, USA

https://doi.org/10.1007/s00125-023-05877-9 / Published online: 10 February 2023

Abbreviations

AC1 Gwet’s coefficient of inter-rater agreement reliability

ADAP Antibody-dependent agglutination PCR

APPA Average pairwise per cent agreement

AS95 Adjusted sensitivity at 95% specificity

AS99 Adjusted sensitivity at 99% specificity

AS100 Adjusted sensitivity at 100% specificity

CLIA Chemiluminescence immunoassay

DASP Diabetes Autoantibody Standardization Program

DK units National Institute of Diabetes and

Digestive and Kidney Diseases consortium units

ECL Electrochemiluminescence

FDR First-degree relative

FloCMIA Flow cytometry microsphere immunoassay

IAS Insulin autoimmune syndrome

IASP Islet Autoantibody Standardization Program

IDS Immunology of Diabetes Society k

Fleiss’ coefficient of inter-rater agreement reliability

LBI Luminex bead immunoassay

LCSP Lowest Concentration Scored Positive

LIPS Luciferase immunoprecipitation system

mAb Monoclonal antibody

OCCC Overall concordance correlation coefficient

pAUC95 Partial ROC-AUC at 95% specificity

RBA Radiobinding assay

ROC-AUC Area under the receiver operating

characteristic curve

W Kendall’s ranking agreement coefficient

Introduction

      Antibody reactivity to insulin was first described in individuals undergoing exogenous insulin administration [1] but people with different autoimmune diseases can also produce IAA in the absence of prior treatment with the hormone. IAA were first described in the insulin autoimmune syndrome (IAS, Hirata’s disease) in 1970 [2] and then in type 1 diabetes in 1983 [3]. The development of insulin autoimmune hypoglycaemia has been linked to exposure to environmental triggers (e.g. drugs and food supplements) in individuals with IAS [4], while no environmental factors have been confirmed in type 1 diabetes.

      IAA are among the first biomarkers to appear during type 1 diabetes natural history. In asymptomatic individuals at risk of diabetes, the appearance of IAA associates with a faster progression to overt type 1 diabetes and younger age at onset [5, 6]. Additionally, specific features of IAA, such as binding affinity, targeted epitopes and titre, are associated with a higher risk of rapid type 1 diabetes development [7–9]. For these reasons, the measurement of IAA has become a cornerstone of screening strategies for type 1 diabetes [10–13] and a particular focus has been given to the improvement of IAA assays.

      The first attempts at the standardisation of IAA measurement highlighted a large variability of results across centres and assays [14–18]. More recent interlaboratory comparisons studies organised by the Diabetes Autoantibody Standardization Program (DASP) showed only partial improvements in assay performance and results concordance [19, 20].

      The Islet Autoantibody Standardization Program (IASP) has superseded the DASP in promoting the continuous improvement of type 1 diabetes autoantibody assays and disseminating empirically tested best-practice protocols, state-of-the-art reagents and serum standards [21]. The IASP is a collaborative effort supported by the Immunology of Diabetes Society (IDS) and the US NIH, which is run by the University of Florida Pathology Laboratories, Endocrine Autoantibody Laboratory and coordinated by an IDS nominated committee. The IASP pursues its goals through the establishment of a periodic interlaboratory comparison of type 1 diabetes-associated autoantibody measurements, aimed at providing an unbiased assessment of assay performance and improving the concordance of results across laboratories around the world. In IASP workshops the participating laboratories test type 1 diabetes autoantibodies in anonymised type 1 diabetes patient, ‘at-risk’ person, and control serum samples. An unbiased comparison of assay performance is provided through the centralised collection and analysis of results by the IASP committee.

       In this report, we present the results of the 2018 and 2020 IASP IAA assays interlaboratory comparison studies that were preliminarily presented at the IASP 2018 and IASP 2020 workshops, held at the 16th and 17th IDS Congress, respectively.

Methods

Study design

      The study was aimed at comparing assay performance across laboratories. Participating laboratories received the same sera in anonymised sets, each labelled with an aliquot-specific unique code. Sera were obtained from the following individuals: individuals with new-onset type 1 diabetes (contributed by several centres around the world), collected within 14 days of the first insulin treatment; multiple islet autoantibodypositive first-degree relatives (FDR) of individuals with type 1 diabetes (enrolled in the TrialNet Ancillary Study – Pathway to Prevention and showing a transiently altered GTT during screening); blood donors without diabetes, collected in the USA. Due to their heterogeneous origin and to the difficulty of procuring sufficiently large serum volume from young children, these samples were only partially representative of incident new-onset diabetes and age-matched control individuals.

      All samples were collected upon written informed consent, with the approval of local ethics committees and according to the ethical principles for medical research involving human subjects of the Declaration of Helsinki [22].

      In 2018, the set included samples from 43 individuals with new-onset diabetes, seven multiple islet autoantibody-positive FDRs, 90 blood donor control individuals, six serial dilutions (156, 20, 5, 1.2, 0.6 and 0 ng/ml) in normal human serum of HUI-018, an IgG1 anti-insulin mouse monoclonal antibody (mAb) targeting a conformational epitope spanning the insulin A and B chains [23], and four additional samples from nondiseased individuals with type 1 diabetes autoantibodies.

      Individuals with type 1 diabetes had a median age of 14 years (range 8–47) and included 15 female individuals and 28 male individuals; 37 were White, two Black, two of mixed ancestry and two of undisclosed ancestry. The FDRs had a median age of 16 years (range 12–53) and included four female and three male individuals, all of White ancestry. Demographic data were available only for 88 non-diabetic blood donors, who had a median age of 20 years (range 18–30), and included 44 female and 44 male individuals, of whom 69 were White and 19 Black.

       In 2020, the set included samples from 38 individuals with new-onset diabetes, 12 multiple islet autoantibody-positive FDRs, 90 control blood donors, HUI-018 mAb serial dilutions, and four standards from the National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK) consortium, corresponding to 235, 5.8, 2 and 0 DK units, made from a mixture of GADA- and IA-2A-positive sera and used as reference material to harmonise antibody assays [24].

      Individuals with type 1 diabetes had a median age of 14 years (range 8–47) and included 15 female and 23 male individuals, of whom 31 were White, two Hispanic, three Black, one of mixed ancestry and one of undisclosed ancestry. FDRs had a median age of 18 years (range 10–53) and included seven female and five male individuals, of whom 11 were White and one was of mixed ancestry.

      Only a minority of samples were present in both 2018 and 2020 sample sets (21 type 1 diabetes, six FDRs, one blood donor) (electronic supplementary material [ESM] Table 1).

Data analysis

      Laboratory personnel were asked to report details of their assay protocol, assay raw data and results using uniform Excel reporting sheets. All data analyses were performed in the R language and environment for statistical computing and graphics [25].

      Assays sensitivity and specificity was calculated as the percentage of case (new-onset plus FDRs) sera reported as IAA positive and as the percentage of blood donor sera reported as IAA-negative, respectively. Adjusted sensitivity at 95% (AS95), 99% (AS99) and 100% specificity (AS100) were calculated after placing the threshold for positivity at the 95th, 99th and 100th percentiles of values observed in the blood donor samples in each assay, respectively.

      Concordance of laboratory-assigned positive or negative scores across assays was expressed as average pairwise per cent agreement (APPA) between assays (i.e. the average number of times each possible combination of two assays agreed on IAA-positive/negative scores divided by the number of samples scored). We tested the occurrence of agreement by pure chance by calculating Gwet’s coefficient of inter-rater agreement reliability (AC1) [26] and Fleiss’ coefficient of inter-rater agreement reliability (k) [27] using the irr R package [28].

      Assay performance in discriminating case from control samples was analysed using the area under the receiver operating characteristic curve (ROC-AUC) and the partial ROCAUC at 95% specificity (pAUC95) [29] using the pROC R package [30].

      Interassay antibody titre concordance was analysed by calculating Kendall’s ranking agreement coefficient (W) [31], after ranking of case and control samples according to autoantibody levels in each assay. The significance of differences in mean ranking of selected case samples between different assays was tested using the Mann–Whitney test.

      In a subset of assays with good performance (pAUC95 ≥0.015) and an immune-complex capture system compatible with the measurement of an anti-mouse IgG mAb, local units were converted into common units (HUI-018 ng/ml equivalents) by applying a linear model to the local units attributed to the provided HUI-018 dilutions followed by rescaling of local quantitative results. The concordance of antibody units was then evaluated by calculating the overall concordance correlation coefficient (OCCC) according to Barnhart using the epiR R package [32].

      For all statistical analyses, two-tailed p values <0.05 were considered as significant.

Results

Summary of submitted IAA assay formats in the IASP 2018 and IASP 2020 workshops

      In the IASP 2018 workshop, 23 laboratories from 13 countries submitted results from 45 IAA assays. The breakdown of assays according to format was as follows: radiobinding assay (RBA) [33]; antibody-dependent agglutination PCR (ADAP) [34]; luciferase immunoprecipitation system (LIPS) [35]; electrochemiluminescence (ECL) [36]; chemiluminescence immunoassay (CLIA); ELISA assay (of which some measured antibodies to oxidised insulin [37]); and Luminex bead immunoassay (LBI) (Fig. 1).

      All RBAs used recombinant insulin radiolabelled with 125I. LIPS assays used proinsulin (n=5) or insulin antigens (n=4) with the NanoLuc luciferase reporter tagged either at the Cterminus of the B chain ([pro]insulin-B-NLuc) or the Nterminus of the A chain (insulin-A-NLuc). Most RBA and LIPS assays were competitive assays performed with or without the presence of untagged insulin competitor (5.6 × 10−6 mol/l and 3.6 × 10−7 mol/l, respectively).

      Antigen–antibody binding occurred in liquid phase in 35 assays (ADAP, ECL, LIPS, RBA) and was followed by the capture of immune complexes via recovery of immunoglobulins (LIPS, RBA) or tagged antigen (ADAP, ECL). Antigen–antibody binding occurred in hybrid solid/liquid phase in two assays (CLIA) and in solid phase for eight (ELISA, LBI). Major characteristics and metrics of each individual assay are reported in ESM Table 2.

      In the IASP 2020 workshop, 22 laboratories from 11 countries submitted results from 37 IAA assays (Fig. 2) based on the following formats: RBA; ADAP; LIPS; ECL; CLIA; and Flow cytometric microsphere-based immunoassay (FloCMIA) [38].

      All RBAs used recombinant insulin radiolabelled with 125I. LIPS assays used either proinsulin (n=6) or insulin (n=5) antigens tagged with a NanoLuc luciferase reporter at the Cterminus of the insulin B chain. One LIPS assay (Duplex LIPS) combined individual measurement of IA-2A and IAA using a dual luciferase system. One laboratory used two different concentrations of competitor in LIPS (3.6 × 10−7 mol/l and 1.1 × 10−9 mol/l).

      Antigen–antibody binding occurred in liquid phase in 36 assays (ADAP, ECL, FloCMIA, LIPS, RBA) or in hybrid solid/liquid phase (CLIA) and was followed by the capture of immune complexes through the recovery of immunoglobulins (LIPS, RBA) or antigens with different tags (ADAP, ECL, FloCMIA, CLIA). Major characteristics and metrics of each individual assay are reported in ESM Table 3.

Assay sensitivity and specificity based on laboratoryassigned scores in the IASP 2018 and IASP 2020 workshops

      In the IAA assays submitted to the IASP 2018 workshop, laboratory-assigned scores showed a median assay sensitivity of 32.0% (IQR 16.0–46.0) and a specificity of 96.7% (IQR 89.7–97.8), with a wide range for both (sensitivity 66.0– 2.0%; specificity 100.0–45.6%) (Fig. 1, ESM Table 2 and ESM Fig. 1).

      In the assays submitted to the IASP 2020 workshop, the median assay sensitivity was 46.0% (IQR 26.0–56.0) and the specificity 98.9% (IQR 96.7–100.0), with a wide range of both (sensitivity 66.0–2.0%; specificity 100.0–80.0%) (Fig. 2, ESM Table 3 and ESM Fig. 2).

ROC-AUC analysis of assay format performance in the IASP 2018 and IASP 2020 workshops

      We evaluated the performance of the IAA assays using the full ROC-AUC and the partial ROC-AUC after imposing a specificity of ≥95%, as a more relevant proxy of assay performance [29] (Fig. 3).

      In the IASP 2018 workshop, IAA assays showed a median ROC-AUC of 0.736 (IQR 0.617–0.803, range 0.232–0.874) and a median pAUC95 of 0.016 (IQR 0.004–0.021, range 0–0.032), against a theoretical pAUC95 maximum of 0.05 (Figs 1, 4 and ESM Fig. 3a). A wide heterogeneity of performance was present both within and across assay

      In the IASP 2020 assays, the median ROC-AUC of the IAA assays was 0.790 (IQR 0.730–0.836, range 0.378– 0.924), while the median pAUC95 was 0.023 (IQR 0.014– 0.026, range 0–0.032) (Figs 2, 4 and ESM Fig. 3b). Additionally, in the IASP 2020 workshop, assay performance varied widely both within and across formats.

Qualitative concordance of laboratory-assigned positive/negative scores

      In the assays submitted to the IASP 2018 workshop, the APPA of positive/negative scores assigned to type 1 diabetes cases across all assays of IAA was 67.1%, while the first-order AC1 of these scores was 0.415. The concordance analysis was greater in control samples (APPA 86.5%, AC1 0.863). When the analysis was limited to assays with a pAUC95 above the median (pAUC95 ≥0.013), these concordance variables improved in both case (APPA 73.4%; AC1 0.476) and control samples (APPA 96.1%; AC1 0.959).

      In the assays submitted to the IASP 2020 workshop, in cases the APPA was 66.9% and the AC1 0.360, while in controls the respective values were 93.7% and 0.932. In assays with a pAUC95 above the median (pAUC95 ≥0.023), these values improved in both case (APPA 74.8%; AC1 0.497) and control samples (APPA 97.3%; AC1 0.972).

      In both years, the observed low Fleiss’ k concordance coefficients were consistent, with most of the control samples scored IAA positive being sporadically so in only a small fraction of assays.

      Overall, the concordance of IAA-positive/-negative scores was greater across assays using the same format (ESM Tables 4–7).

Assay format-specific patterns of IAA recognition

      In both IASP workshops, discrepancies in positive/negative scores showed format-specific patterns. In 2018, a subset of type 1 diabetes cases (IDS326, IDS322, IDS292, IDS298, IDS004, IDS309, IDS006, IDS290) was IAA positive predominantly in ADAP and LIPS and in few RBA and ECL assays. Conversely, a different subset of sera (IDS303, IDS301) was IAA positive in ADAP and most RBAs but only in a minority of LIPS and ECL assays. A subset of control sera (TS24176, N59807, N54153) was IAA positive mostly in LIPS and sporadically in other formats, while another control sample (C1401) was positive mostly in RBA and ELISA (Figs 5, 6 and ESM Figs 4, 5).

     In the sera submitted to the IASP 2020 workshop, two case samples (IDS359, IDS326) were IAA positive predominantly in LIPS but only in a minority of RBA and ECL assays. Conversely, another subset of case samples (IDS324, IDS303 and IDS351) was positive in most RBA and ECL but only in a minority of LIPS assays. Among control sera, two (S6320, S8768) were positive mostly in LIPS but only sporadically in other formats, while two other serum samples (S6389, LQ22722) were positive exclusively in ECL. In LIPS assays using two different concentrations of unlabelled insulin competitor, IAA positivity was sometimes conditional upon the use of a high concentration of competitor (IDS326), suggesting the presence of low-affinity IAA. Always in LIPS, one serum sample (IDS351) resulted positive only when insulin antigen was used, suggesting that antigen recognition was conditional on removal of the C-peptide (Figs 7, 8 and ESM Figs 6, 7).

      The results also highlighted the selective recognition of some case sera by assays in which antibody–antigen binding occurs in liquid phase (ADAP, ECL, LIPS, RBA) vs solid phase (CLIA, ELISA, FloCMIA, LBI) in both IASP 2018 (IDS312, IDS009, IDS268, IDS328, IDS334, IDS317) and IASP 2020 (IDS372, IDS345, IDS317, IDS312, IDS365, IDS310, IDS358, IDS367) workshops.

Ranking of autoantibody levels

      Quantitative interassay concordance of IAA levels was evaluated by ranking sera in each assay and then calculating the Kendall’s W ranks agreement coefficient.

      In the IASP 2018 workshop, the W coefficient across all assays was 0.354 for case sera and 0.081 for control sera. In the IASP 2020 workshop, the W coefficient across all assays was 0.458 for case sera and 0.047 for control sera.

      In both workshops, excluding assays with low performance (i.e. low pAUC95) from the analysis led to a modest increase of the agreement coefficient in both case and control samples (IASP 2018, W=0.431 and 0.086, respectively; IASP 2020, W=0.587 and 0.057, respectively). Limiting the analysis to assays with higher performance (i.e. greater than the median pAUC95) showed a further increase of W for case sera but only a marginal improvement for control sera (IASP 2018, W=0.656 and 0.098, respectively; IASP 2020, W=0.617 and 0.086, respectively).

      Concordance of IAA level ranks increased among assays using the same format for both case and control sera (ESM Figs 8–15 and ESM Tables 4–7). The comparison of assay formats’ W coefficients showed that ranking agreement was higher in LIPS than in both local and commercial RBAs, CLIA and ECL assays.

      IAA level ranking also confirmed the previously observed preferential recognition of some case samples by assays in which antibody–antigen binding occurs in liquid phase vs solid phase (Mann–Whitney test, all p≤0.001 in IASP 2018 workshop and all p<0.05 in IASP 2020 workshop).

Autoantibody levels in 2018 vs 2020

      Using the 28 samples that were distributed in both the IASP 2018 workshop and the IASP 2020 workshop, we analysed the correlation of quantitative results in 24 assays (nine local RBAs, one commercial RBA, one ADAP, five LIPS, one CLIA and seven ECL). Correlation was highest in RBA, LIPS and ADAP (median R2 =0.97 [IQR 0.94–0.98]) and lower in ECL and CLIA (median R2 =0.25 [IQR 0.01–0.50]) (ESM Figs 16, 17).

Conversion of local units into common HUI-018 units

      Using the provided HUI-018 anti-insulin mAb serial dilutions, we converted local arbitrary units into common units (HUI- 018 ng/ml) by applying a log–log linear regression model. In both the IASP 2018 and the IASP 2020 workshop, the correlation of local and common units was high (R2 range 0.85– 1.00) but the slopes of the regression curves varied considerably across assays (slope range 0.406–1.000) (ESM Figs 18, 19). We then compared calculated common HUI-018 units in each assay with the true concentration of the HUI-018 serial dilutions (ESM Figs 20, 21). The observed high degree of variability of the assigned concentrations was suggestive of potentially large discrepancies in the linear range of the different assays.

Autoantibody levels in common HUI-018 units

      We converted local arbitrary units into common HUI-018 units for assays with good performance and a detection system compatible with the measurement of a mouse IgG mAb (Fig. 9). We then calculated the OCCC of HUI-018 units as another measure of interassay quantitative concordance. In both workshops, the OCCC was relatively low in both case (IASP 2018, 0.285; IASP 2020, 0.203) and control sera assays (IASP 2018, 0.013; IASP 2020, 0.012). After stratification of assays according to format and antigen, the quantitative concordance of common units improved only for some assays (ESM Figs 22, 23 and ESM Tables 4–7).

Analytical sensitivity of IAA assays

      As a proxy of the IAA assays’ analytical sensitivity, we determined the lowest concentration scored positive (LCSP) by each assay for HUI-018 mAb (Figs 1, 2, 5, 7). The median LCSP was 1.2 ng/ml in both workshops but exhibited extreme variability across assays (range 0.6–156). In addition, some assays recognised the diluent normal serum as weakly positive.

Discussion

      Early interlaboratory comparison studies demonstrated that the detection of disease-specific IAA was crucially dependent on assay format choice and led to the establishment of the liquid-phase immunoprecipitation RBA assay as the de facto gold standard for IAA measurement [14–20].

      The more recent IASP 2018 and IASP 2020 interlaboratory IAA measurement comparison studies saw not only the continued implementation of RBAs by most laboratories but also an increasingly wider adoption of alternative nonradioactive formats. Therefore, in this paper we were able to perform a comprehensive comparison of RBAs and other assay formats in terms of diagnostic performance, sensitivity, specificity and concordance.

      RBAs submitted to the IASP workshop in both 2018 and 2020 were predominantly in-house assays derived from the IAA micro-assay originally described in 1997 by Williams et al [33]. The performance of these micro-assays varied widely in both workshops but compared with more recent non-radioactive IAA immunoassays most RBAs showed similar or better sensitivity and specificity, as well as pAUC95 and AS95, two metrics aimed at excluding regions of low assay specificity and poor clinical relevance from the analysis. Of note, commercial RBAs, while showing good specificity, had a sensitivity less than half that of most in house micro-assays, possibly because of a lower capacity to immuno-precipitate IAA immune complexes by the antihuman IgG polyclonal antibody used in commercial kits compared with the protein A/G coated Sepharose beads adopted by the micro-assay RBA.

      Among non-radioactive IAA immunoassays, the only submitted ADAP assay, the most recently developed IAA assay format, showed the highest or second-highest pAUC95 and AS95.

      In both workshops, among non-radioactive assay formats, the most widely adopted were LIPS and ECL. Submitted LIPS assays implemented a variety of alternative protocols and antigens and showed variable performance. Nevertheless, LIPS assays showed the overall second-highest median pAUC95 and AS95 after RBA.

      ECL assays comprised two major variants, the first measuring IAA potentially of any Ig class (pan-Ig ECL) and the second multiplexing and discriminating Ig of different classes (IgM, IgA and IgG multiplexed-ECL). In both 2018 and 2020, pan-Ig ECL assays showed a relatively good performance, although lower than that of the best RBAs. Multiplexed IgG, IgM and IgA ECL assays instead were not only less sensitive, as might have been expected, but also less specific, as signalled by the scoring as IgM and/or IgA IAA positive of some of the included dilutions of HUI-018, a mouse antiinsulin mAb of the IgG1 class.

     The remaining non-radioactive assays submitted to the workshops comprised a variety of formats such as CLIA, FloCMIA, LBI, ELISAs, and ELISAs using oxidised insulin as antigen. All of these assays showed a poor performance, with drastically reduced sensitivity and specificity compared with RBA and their ROC-AUCs demonstrated an inability to discriminate case sera from control sera.

      In both workshops, we observed discrepancies of positive/ negative scores and ranking of antibody levels across assays and formats, even when the analysis was limited to assays with overall good performance. The underlying reasons for the selective recognition of some case and control samples as IAA positive in certain assay formats but not others remain to be fully clarified. However, two main mechanisms can be put forward to explain these discrepancies: the first is simply linked to the difficulty of lower performance assays in identifying as positive sera presenting with low level IAA; and the second presumes a differential recognition of insulin epitopes by antibodies present only in a subset of sera. Multiple causes might underly this potential second mechanism, such as possible alteration of some insulin epitopes in certain assay formats caused by the addition of tags (e.g. biotin residues in ECL or luciferase enzyme in LIPS) or by potential alternative antigen post-translational modification(s) in the different expression systems used for their production. This potential formatassociated selective recognition of some epitopes remains to be evaluated but might have an important impact on autoantibody-based screening strategies, which currently are still based on RBA.

      In both the IASP 2018 and IASP 2020 workshops, the comparison of quantitative IAA results was complicated by the variety of local non-standardised arbitrary units and calculation algorithms into which results were expressed. In the absence of a WHO-recognised IAA standard serum, we explored the possibility of using an anti-insulin mouse monoclonal antibody as reference. For this reason, we distributed anonymised serial dilutions of the HUI-018 mAb. Most liquid-phase assays showed a clear ability to detect HUI-018 binding to insulin but the correct ranking of the mAb dilutions was challenging for assays with lower performance. While neither IASP study was designed to determine rigorously the analytical sensitivity of participating assays, using the differential recognition of the HUI-018 mAb dilution as a proxy, we could infer the presence of important differences in analytical sensitivity across IAA assays.

     Moreover, the conversion of local units into common HUI-018 units further confirmed the lack of good quantitative concordance of IAA assays even when using the same format, a phenomenon that could not be conclusively clarified within the limits of the current study design.

      Legislative and logistic pressure against the use of radioactive substances spur the development and validation of novel non-radioactive immunoassays. Furthermore, the expected future implementation of antibody-based population screening programmes for type 1 diabetes would benefit from the implementation of high-performance IAA assays dispensing with the need for radio-isotopic tracers. In this context, while in-house micro-assay RBAs still constituted the majority of the best-performing assays in 2020, some non-radioactive formats could indeed achieve both high sensitivity and specificity (e.g. ADAP, LIPS and ECL). However, none of the classical immunoassay formats widely adopted in routine clinical diagnostics (e.g. ELISA, CLIA and Luminex bead-based assays) nor assays aimed at measuring Ig class-specific autoantibody responses can be currently recommended for IAA measurement in light of their poor demonstrated performance.

      In conclusion, our research supports the value of type 1 diabetes autoantibody assay evaluation programmes. These programmes not only help to assess the accuracy of diagnostic tests objectively but also provide academic research laboratories and companies with the chance to learn and improve their immunoassays.

Supplementary Information The online version contains peer-reviewed but unedited supplementary material available at https://doi.org/10.1007/ s00125-023-05877-9.

Acknowledgements We wish to thank all the people and the clinicians who contributed blood samples, without whose generosity and effort the IASP workshops and collaborative studies could not take place. We also wish to remember the precious contribution to both studies’ conception and data analysis of our friend and IASP committee member Alistair J. K. Williams, who passed away in 2020.

Data availability The datasets generated and analysed during the current study are available from the corresponding author on reasonable request and provided the permission of all participating laboratories is given.

Funding Open Access funding enabled and organized by Projekt DEAL. The authors acknowledge the support received by the IASP study through a TEDDY [39] grant administered at the University of South Florida (Tampa, FL, USA) and by the Type 1 Diabetes TrialNet/IASP Ancillary studies arm [40].

Authors’ relationships and activities PA is a member of the Editorial Board of Diabetologia. All other authors declare that there are no relationships or activities that might bias, or be perceived to bias, their work.

Contribution statement

AJW, PA, MS, BA, DLP, WEW and VL contributed to the conception and design of the study. DLP and WEW organised and supervised sample preparation and distribution and data collection from participating laboratories. IM, DLP and VL contributed to the collection, preliminary analysis and interpretation of data. IM and VL drafted the initial manuscript. AJW, AEL, PA, MS, BA, DLP and WEW critically revised the data analysis and the manuscript. All authors approve of the publication. VL is the guarantor of the paper and accepts full responsibility for the work and/or the conduct of the study, had access to the data, and controlled the decision to publish.

Participating laboratories and contacts for the IASP 2018 and IASP 2020 IAA interlaboratory comparison studies

S. N. Valdez, Facultad de Farmacia y Bioquímica, Universidad de Buenos Aires, Buenos Aires, Argentina;

K. Watson, Royal Melbourne Hospital, Melbourne, VIC, Australia;

K. Verhaeghen, Uzbrussel Vub, Clinical Biology of Diabetes

Diabetes Research Center, Brussels, Belgium;

V. Chen, Laboratory of Snibe, Shenzhen, China;

J. Zhang, Shenzhen YHLO Biotech Co., Ltd, Shenzhen, China;

Z. Zhou, G. Huang, Diabetes Center, Central South University, Changsha, China;

R. Uibo, K. Reimand, University of Tartu, Department of Immunology, Tartu, Estonia;

M. Knip, T. Härkönen, Children’s Hospital, Scientific Laboratory, University of Helsinki, Helsinki, Finland;

R. Veijola, Department of Pediatrics, Diabetes Research Laboratory, University of Oulu, Oulu, Finland;

L. Chatenoud, Laboratoire d’Immunologie Biologique-Hôpital Necker-Enfants Malades Paris, Paris, France;

D. Mueller, Preclinical Approaches to Stem Cell Therapy/ Diabetes, Dresden, Germany;

P. Achenbach, Institute of Diabetes Research, Helmholtz Zentrum München, Neuherberg, Germany;

M. Schlosser, University Medical Center Greifswald, Karlsburg,Germany;

V. Lampasona, Diabetes Research Institute, IRCCS Istituto Scientifico San Raffaele, Milano, Italy

B. Almås, The Hormone Laboratory, Haukeland University Hospital, Bergen, Norway;

K. S. Opsion, Hormone Laboratory, Oslo University Hospital, Oslo, Norway;

A. Ramelius, Diabetes And Celiac Disease Research Unit, Lund University, Malmö, Sweden;

I. Johansson, Clinical & Experimental Research, Division of Pediatrics, Linköping, Sweden;

M. R. Batstra, T. Cieremans, Reinier De Graaf Groep, Department of Medical Immunology, Delft, the Netherlands;

S. Kostense, World Without Disease, Janssen, Leiden, the Netherlands;

A. E. Long, Diabetes & Metabolism, Learning & Research, University of Bristol, Bristol, UK;

A. Mathew, Meso Scale Diagnostics, LLC., Rockville, MD, USA;

C. Hampe, University of Washington, Seattle, WA, USA;

C. Lu, Meso Scale Discovery, Gaithersburg, MD, USA;

C. Mann, Quest Diagnostics Nichols Institute, San Juan Capistrano,CA, USA;

L. Yu, Barbara Davis Center, Aurora, CO, USA;

M. Mamula, L2 Diagnostics at Yale University, New Haven, CT, USA;

P. Robinson, Enable Biosciences, INC, San Francisco, CA, USA;

W. A. Hagopian, Pacific Northwest Diabetes Research Institute,Seattle, WA, USA.

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ .

References

1. Moinat P, Marston E (1958) A quantitative estimation of antibodies to exogenous insulin in diabetic subjects. Diabetes 7(6):462–467. https://doi.org/10.2337/diab.7.6.462

2. Cappellani D, Macchia E, Falorni A, Marchetti P (2020) Insulin autoimmune syndrome (Hirata disease): a comprehensive review fifty years after its first description. Diabetes Metab Syndr Obes Targets Ther 13:963–978. https://doi.org/10.2147/DMSO.S219438

3. Palmer JP, Asplin CM, Clemons P et al (1983) Insulin antibodies in insulin-dependent diabetics before insulin treatment. Science 222(4630):1337–1339. https://doi.org/10.1126/science.6362005

4. Censi S, Mian C, Betterle C (2018) Insulin autoimmune syndrome: from diagnosis to clinical management. Ann Transl Med 6(17): 335–335. https://doi.org/10.21037/atm.2018.07.32

5. Vehik K, Bonifacio E, Lernmark Å et al (2020) Hierarchical order of distinct autoantibody spreading and progression to type 1 diabetes in the TEDDY study. Diabetes Care 43(9):2066–2073. https:// doi.org/10.2337/dc19-2547

6. Krischer JP, Liu X, Lernmark Å et al (2021) Characteristics of children diagnosed with type 1 diabetes before vs after 6 years of age in the TEDDY cohort study. Diabetologia 64(10):2247–2257. https://doi.org/10.1007/s00125-021-05514-3

7. Achenbach P, Koczwara K, Knopff A, Naserke H (2004) Mature high-affinity immune responses to (pro) insulin anticipate the autoimmune cascade that leads to type 1 diabetes. J Clin Invest 114(4): 589–597. https://doi.org/10.1172/JCI200421307

8. Schlosser M, Koczwara K, Kenk H et al (2005) In insulinautoantibody-positive children from the general population, antibody affinity identifies those at high and low risk. Diabetologia 48(9):1830–1832. https://doi.org/10.1007/s00125-005-1864-6

9. Achenbach P, Schlosser M, Williams AJK et al (2007) Combined testing of antibody titer and affinity improves insulin autoantibody measurement: Diabetes Antibody Standardization Program. Clin Immunol 122(1):85–90. https://doi.org/10.1016/j.clim.2006.09. 004

10. Elding Larsson H, Vehik K, Gesualdo P et al (2014) Children followed in the TEDDY study are diagnosed with type 1 diabetes at an early stage of disease: Intense follow-up enables early detection of type 1 diabetes. Pediatr Diabetes 15(2):118–126. https://doi. org/10.1111/pedi.12066

11. Krischer JP, Liu X, Vehik K et al (2019) Predicting islet cell autoimmunity and type 1 diabetes: an 8-year TEDDY study progress report. Diabetes Care 42(6):1051–1060. https://doi.org/10.2337/ dc18-2282

12. Ziegler A-G, Kick K, Bonifacio E et al (2020) Yield of a public health screening of children for islet autoantibodies in Bavaria, Germany. JAMA 323(4):339. https://doi.org/10.1001/jama.2019. 21565

13. Karpen SR, Dunne JL, Frohnert BI et al (2022) Consortium-based approach to receiving an EMA qualification opinion on the use of islet autoantibodies as enrichment biomarkers in type 1 diabetes clinical studies. Diabetologia. https://doi.org/10.1007/s00125-022- 05751-0

14. Wilkin T, Palmer J, Bonifacio E et al (1987) First international workshop on the standardisation of insulin autoantibodies. Diabetologia 30(8):676–677. https://doi.org/10.1007/BF00277328

15. Wilkin T, Palmer J, Kurtz A, Bonifacio E, Diaz JL (1988) The second international workshop on the standardisation of insulin autoantibody (IAA) measurement. Held in New York, 27-30 October 1987. Diabetologia 31(7):449–450. https://doi.org/10. 1007/BF00271590

16. Greenbaum CJ, Palmer JP, Kuglin B, Kolb H (1992) Insulin autoantibodies measured by radioimmunoassay methodology are more related to insulin-dependent diabetes mellitus than those measured by enzyme-linked immunosorbent assay: results of the Fourth International Workshop on the Standardization of Insulin. J Clin Endocrinol Metab 74(5):1040–1044. https://doi.org/10.1210/jcem. 74.5.1569152

17. Greenbaum CJ, Wilkin TJ, Palmer JP (1992) Fifth international serum exchange workshop for insulin autoantibody (IAA) standardization. The immunology and diabetes workshops and participating laboratories. Diabetologia 35(8):798–800. https://doi.org/10. 1007/BF00429105

18. Wilkin TJ, Schoenfeld SL, Diaz JL, Kruse V, Bonifacio E, Palmer JP (1989) Systematic variation and differences in insulinautoantibody measurements. Diabetes 38(2):172–181. https://doi. org/10.2337/diabetes.38.2.172

19. Bingley PJ, Bonifacio E, Mueller PW (2003) Diabetes antibody standardization program: first assay proficiency evaluation. Diabetes 52(5):1128–1136. https://doi.org/10.2337/diabetes.52.5. 1128

20. Schlosser M, Mueller PW, Törn C, Bonifacio E, Bingley PJ, and participating laboratories (2010) Diabetes antibody standardization program: evaluation of assays for insulin autoantibodies. Diabetologia 53(12):2611–2620. https://doi.org/10.1007/s00125- 010-1915-5

21. Lampasona V, Pittman DL, Williams AJ et al (2019) Islet autoantibody standardization program 2018 workshop: interlaboratory comparison of glutamic acid decarboxylase autoantibody assay performance. Clin Chem 65(9):1141–1152. https://doi.org/10. 1373/clinchem.2019.304196

22. World Medical Association (2001) World Medical Association declaration of Helsinki. Ethical principles for medical research involving human subjects. Bull World Health Organ 79(4):373–374

23. Johansson E, Wu X, Yu B et al (2021) Insulin binding to the analytical antibody sandwich pair OXI-005 and HUI-018: Epitope mapping and binding properties. Protein Sci 30(2):485–496. https://doi.org/10.1002/pro.4009

24. Bonifacio E, Yu L, Williams AK et al (2010) Harmonization of glutamic acid decarboxylase and islet antigen-2 autoantibody assays for national institute of diabetes and digestive and kidney diseases consortia. J Clin Endocrinol Metab 95(7):3360–3367. https://doi.org/10.1210/jc.2010-0293

25. R Core Team (2020) R: A Language and Environment for Statistical Computing. R Foundation for Statistical Computing, Vienna, Austria

26. Wongpakaran N, Wongpakaran T, Wedding D, Gwet KL (2013) A comparison of Cohen’s Kappa and Gwet’s AC1 when calculating inter-rater reliability coefficients: a study conducted with personality disorder samples. BMC Med Res Methodol 13(1):61. https:// doi.org/10.1186/1471-2288-13-61

27. Cicchetti DV, Feinstein AR (1990) High agreement but low kappa: II. Resolving the paradoxes. J Clin Epidemiol 43(6):551–558. https://doi.org/10.1016/0895-4356(90)90159-M

28. Gamer M, Lemon J, Fellows I, Singh P (2019) irr: Various Coefficients of Interrater Reliability and Agreement. R package version 0.84.1. Available from: https://CRAN.R-project.org/ package=irr (accessed on 26 January 2019)

29. Ma H, Bandos AI, Rockette HE, Gur D (2013) On use of partial area under the ROC curve for evaluation of diagnostic performance. Stat Med 32(20):3449–3458. https://doi.org/10.1002/sim.5777

30. Robin X, Turck N, Hainard A et al (2011) pROC: an open-source package for R and S+ to analyze and compare ROC curves. BMC Bioinformatics 12(1):77. https://doi.org/10.1186/1471-2105-12-77

31. Kendall MG, Babington Smith B (1939) The problem of m rankings. Ann Math Stat 10(3):275–287

32. Stevenson M, Sergeant E, Nunes T et al (2021) epiR: tools for the analysis of epidemiological data. R package version 2.0.26. https:// CRAN.R-project.org/package=epiR (accessed on 15 July 2021)

33. Williams AJ, Bingley PJ, Bonifacio E, Palmer JP, Gale EA (1997) A novel micro-assay for insulin autoantibodies. J Autoimmun 10(5):473–478. https://doi.org/10.1006/jaut.1997.0154

34. Tsai C, Robinson PV, Spencer CA, Bertozzi CR (2016) Ultrasensitive antibody detection by agglutination-PCR (ADAP). ACS Cent Sci 2(3):139–147. https://doi.org/10.1021/acscentsci. 5b00340

35. Liberati D, Wyatt RC, Brigatti C et al (2018) A novel LIPS assay for insulin autoantibodies. Acta Diabetol 55(3):263–270. https:// doi.org/10.1007/s00592-017-1082-y

36. Yu L, Dong F, Miao D, Fouts AR, Wenzlau JM, Steck AK (2013) Proinsulin/insulin autoantibodies measured with electrochemiluminescent assay are the earliest indicator of prediabetic islet autoimmunity. Diabetes Care 36(8):2266–2270. https:// doi.org/10.2337/dc12-2245

37. Strollo R, Vinci C, Napoli N, Pozzilli P, Ludvigsson J, Nissim A (2017) Antibodies to post-translationally modified insulin as a novel biomarker for prediction of type 1 diabetes in children. Diabetologia 60(8):1467–1474. https://doi.org/10.1007/s00125- 017-4296-1

38. Sabljic AV, Bombicino SS, Marfía JI et al (2021) Novel flow cytometric immunoassay for detection of proinsulin autoantibodies in diabetes mellitus employing a recombinant autoantigen expressed in E. coli. Front Immunol 12:648021. https://doi.org/10.3389/ fimmu.2021.648021

39. TEDDY Study Group (2008) The environmental determinants of diabetes in the young (TEDDY) study. Ann N Y Acad Sci 1150:1–13. https://doi.org/10.1196/annals.1447.062

40. Skyler JS, Greenbaum CJ, Lachin JM et al (2008) Type 1 Diabetes TrialNet–an international collaborative clinical trials network. Ann N Y Acad Sci 1150:14–24. https://doi.org/10.1196/annals.1447.054

Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

This article is excerpted from the Diabetologia (2023) 66:897–912 by Wound World.