Introduction
Advanced hybrid closed-loop (AHCL) systems represent the state of art of insulin therapy in people with type 1 dia-betes (T1D). They combine three components: a continuous subcutaneous insulin pump, a sensor for continuous glucose monitoring (CGM), and a control algorithm that constantly analyzes the sensor-measured interstitial glucose and adjusts the rate of basal insulin accordingly. Manual meal boluses are still required.
Currently, the Italian National Health System allows funded access to several AHCL systems approved for adults: the MinimedTM 780G system (Minimed Medtronic, Northridge, California, USA), the Tandem t:slim X2 Control IQTM system (Control-IQ technology, Tandem Inc., San Diego, CA, USA) the Diabeloop Generation 1 (DBLG1® system, Grenoble, France), and the CamAPS FX app (Cam Diab, Cambridge, UK).
These systems have demonstrated in several RCTs [1–3] and real-world studies [4–7] to better achieve and maintain glycemic targets in people with type 1 diabetes compared to multiple dose insulin (MDI) therapy, conventional continu-ous subcutaneous insulin infusion (CSII) or to sensor-aug-mented pump (SAP) with predictive low-glucose suspend systems. Few, most pediatric, studies compared the real-life efficacy of different AHCL systems [8–10]. However, in everyday real life, many confounders can make it difficult to compare different AHCL systems, such as lifestyle dur-ing the day, the precision of the insulin bolus at meals, the degree of physical activity and, last but not least, differences in the overall degree of glycemic control between diabetic subjects.
Precisely for this reason we decided to compare the effec-tiveness of AHCL systems with different control algorithms in a group of patients with similar glycemic control and only during the night-time hours to minimize confounding effects.
Therefore, the aim of this study was to compare the effec-tiveness of three different AHCL systems in adults with T1DM with an almost good glycemic control in achieving the recommended glycemic targets at night-time (00:00 to 07:00).
Materials and methods
We retrospectively evaluated 55 adults with type 1 diabetes mellitus (T1DM) followed by the Unit of Endocrine Dis-eases and Diabetology, ASST Papa Giovanni XXIII (Ber-gamo, Italy), who were already using an AHCL system for at least 6 months.
Inclusion criteria was a similar daily glycemic control with a Glucose Management Index (GMI) between 7.0% and 7.4% in the last 3 months before the night-time com-parison. Furthermore, all subjects had dinner before 9 pm and did not start breakfast before 7am during the 14-day study period.
Twenty-two were using the Medtronic MinimedTM 780G system composed by an insulin pump containing a Predictive Integrative Derivative (PID) algorithm (Smart-Guard algorithm) and by Guardian G4 sensor; 18 were using a system combining an insulin pump Tandem t:slim X2 integrated with a Model Predictive of Control (MPC) algorithm and a Dexcom G6 continuous glucose monitor-ing sensor (Dexcom Inc., San Diego, CA, USA), and 15 were using a system combining an insulin pump Accu-Chek Insight (Roche Diabetes CARE, Basel, Switzerland), a Dex-com G6 continuous glucose monitoring device and a Model Predictive Control algorithm (Diabeloop Generation 1) con-tained in a controller handset.
Demographic, anamnestic, and clinical data of the par-ticipants [i.e., age, gender, weight, height, body mass index (BMI), duration of diabetes], were obtained from the elec-tronic medical record.
Data regarding percentage time spent in hypoglyce-mic (TBR<54 and TBR 54–69 mg/dl), euglycemic (TIR 70–180 mg/dl and tight TIR 70–140 mg/dl), and hyper-glycemic (TAR 181–250 and TAR>250 mg/dl) ranges, CGM-measured mean glucose concentration, glucose man-agement indicator (GMI), coefficient of variation (CV) of CGM-measured glucose concentrations, percentage of sen-sor use, and insulin infusion of 14 consecutive days were downloaded from the platforms Carelink,, Glooko-Diasend and Yourloops.
The participants were categorized in three groups accord-ing to the AHCL system they were using (Minimed 780G group, Control-IQ group and DBLG1 group).
Outcomes
Main outcomes were the differences between the three groups in CGM metrics and in insulin infusion in the night-time period from 00:00 to 07:00.
Statistical analysis
Data are shown as mean±SD.
All p values are two sided. A p value<0.05 was consid-ered significant.
All variables were tested to evaluate whether they were normally distributed using the Kolmogorov–Smirnov test. Depending on the distribution, comparison tests were per-formed using analysis of variance ANOVA (normally dis-tributed data) and Kruskal–Wallis (non-normally distributed data).
To analyze insulin dosage changes over night, we employed a panel data regression model to examine tempo-ral variations. The first time-interval (00:00 to 01:00) served as the reference parameter. We calculated predicted margins to estimate the changes over time and employed a param-eter test to assess the influence of the time of day on insulin dosage.
To assess nocturnal variations in insulin dosage across the three AHCL systems, we compared the administered insulin doses within each hourly interval (00:00 to 01:00, 01:00 to 02:00, …) using the Kruskal–Wallis test.
Analyses were carried out with Stata.
The study was performed in accordance with the Hel-sinki Declaration of 1964 and its later amendments and agreed with national regulations. It was approved by the local ethical committee and written informed consent to use the clinical and biochemical data was obtained from each participant.
Results
Baseline demographic and clinical characteristics of the three groups are shown in Table 1.
Although the subjects in the Diabeloop group and espe-cially those in the Control-IQ group were younger than those in Minimed 780G group, the 3 groups were similar with regard to BMI, age at onset and duration of diabetes, and daily insulin requirement. As per the inclusion criterion, the CGM metrics were comparable in the 3 groups at the time of the 14 days of the study period.
The CGM metrics, the insulin requirement, and the per-centage of sensor and of automatic mode use in the 14 days of the study did not differ between the three groups (Table 1).
Night-time period
The trend of glucose was similar in the three groups with a progressive reduction from 00:00 to 06:00 (Fig. 1A, B and C).
The Control-IQ group spent a significantly (p=0.007) higher percentage of time in tight TIR (51.5±9.8%) com-pared to the Minimed 780G group (42.1±13.7%) and the DBLG1 group (40.1±10.5%) (Table 2).
The CV of CGM-measured glucose concentrations was lower, close to statistical significance (p=0.0507), in the Minimed 780G group (29±6.7%) compared to Control-IQ group (34.5±6.6%) and DBLG1 group (32.4±7.1%).
The other CGM metrics did not differ between the three groups (Table 2).
The insulin requirement did not significantly differ between the three groups, except between 5 and 6 am when it was greater in Diabeloop group (Table 3).
A regression analysis was performed to explore the noc-turnal variation in insulin dosage across different hours of the night in all 55 patients, utilizing a panel data model (Table 4A).
The mixed-effects model revealed significant variability in insulin dosage associated with different hours of the night (Likelihood ratio test p<0.001). The estimated coefficients for each hour represent the change in insulin dosage com-pared to the reference hour (00.00–01.00).
These data suggest that the time of night significantly influences insulin dosage (Fig. 2A illustrate the trend of insulin administration throughout the night in all patients).
The significant negative coefficients associated with hours 02.00 to 06.00 suggest that this time interval is linked to a decrease in the insulin dosage compared to the 00.00–01.00- time interval.
Comparing the total nighttime insulin dosage adminis-tered by the three systems, there are no significant differ-ences. However, upon analyzing specific hours of the night (Table 4), it is noteworthy that DBLG1 group exhibits a sig-nificantly higher insulin dosage administered between 05:00 and 06:00 compared to the other systems (p=0.03), spe-cifically 52% higher than Medtronic and 41% higher than Tandem (Fig. 2 in B, C and D shows the trend of insulin


administration throughout the night across the three aHCL systems).
The hourly coefficient of variation (CV) of exogenous insulin delivery during night-time was calculated to quan-tity intraperson variability of insulin requirements in all participants, and in the 3 groups. The mean of intraperson variability of insulin requirement was high and similar in all hours of night-time period without difference between the three groups (Table 4).
Discussion
To our knowledge this is the first study comparing the effi-cacy of three AHCL systems in optimizing glycemic control over the night-time period.
Minimed 780G group used a PID algorithm unlike group Control-IQ and DBLG1 that used a MPC algorithm.
Recently Kang et al. [11] performed a meta-analysis of randomized controlled trials of three different algorithms (MPC, PID, and Fuzzy) compared to conventional insulin therapy. The MPC algorithm-based artificial pancreas sys-tem was associated with a higher increase in TIR ([MD], 12.57%, 95% CI [9.63, 15.50] p<0.00001) than the PID algorithm-based artificial pancreas system ([MD], 9.59%, 95% CI [−3.67, 22.85] p<0.00001). On the other hand, the reduction in TBR was higher in studies using the PID ([MD], −5.24%, 95% CI [−16.06, 5.58] p<0.00001) algo-rithm than in studies using the MPC ([MD], −1.12%, 95% CI [−1.50 to − 0.75] p<0.00001) algorithm.
In a recent study Bassi et al. [9] retrospectively compared 1-year use of Tandem Control-IQ vs Minimed 780G in 76 adult and pediatric subjects with T1DM. The use of both


AHCL systems led to a significant and rapid improvement of glycemic control lasting for up 12 months. Unlike in Kang’s meta-analysis, subjects using Minimed 780G with a PID algorithm obtained a higher TIR (71% vs 68%; p=0.001) and lower TAR>250 mg/dl (4.5% vs 9%, p=0.009), aver-age sensor glucose (148.5 vs 162 mg/dl, p<0.001), and standard deviation (50 vs 58 mg/dl, p=0.031) than users Tandem Control-IQ system with a MPG algorithm. Simi-larly, a prospective multicenter study, including 150 adoles-cents and adults with type 1 diabetes found that two AHCL (Medtronic 780G and Tandem Control-IQ) were able to provide a similar improvement in glucose control with no superiority of one system over the other [12].
However, to compare properly multiple AHCL systems it is necessary to reproduce similar conditions. As such, we evaluated adult individuals with a similar basal glucose control (HbA1c 53–58 mmol/mol or 7–7.5%): the mean GMI during the study was 55 mmol/mol (7.2%) in all three groups. The overall CGM metrics and the insulin require-ment did not differ in the three group regardless of type of system and algorithm used.
We chose as main outcomes the overnight differences in CGM metrics between the three AHCL systems, because night-time data are more reproducible as there is fewer con-founding external factors, such as food intake and physical activity.
In a randomized, crossover trial study Pinsker and coll. compared the efficacy of a PID vs a MPC algorithm in overnight glucose control after a standard 65 g-dinner in 20 adults with T1DM. Each participant performed two super-vised 27.5 h sessions using a PID and a MPC algorithm. Overnight glucose control was excellent with both algo-rithms, showing similar CGM metrics [13].
Moreover, in our study the overnight glucose control was adequate and the recommended CGM targets were obtained with all three AHCL systems. The Control-IQ group showed a significantly higher percentage of time spent in tight time in range compared to other two groups. It is notable that Control-IQ technology has a “Sleep Activity Mode” to opti-mize glycemic control overnight. If the user activates this mode the algorithm uses a glucose target range of 112.5– 120 mg/dL instead of 112.5–160 mg/dL and increases the


basal insulin infusion when the predicted glucose value is>120 mg/dL [14].
In another retrospective study the use of the Sleep Activ-ity Mode did not lead to an increase in TIR in a pediatric population treated with Tandem–Control-IQ. However, unlike in our study the Authors did not evaluate tight TIR [15].
Sensor glucose values and insulin requirement fol-lowed a similar trend in all three groups, both decreased from midnight to 5 am. After 5 am sensor glucose values were unchanged, but insulin infusion increased displaying a reduction in insulin sensitivity in the period 4–8 a.m. as compared to the period after midnight to 3 a.m. in people with T1DM. Pioneering studies showed a reduction in insu-lin sensitivity in the period 4–8 a.m. in people with T1DM [16, 17].
Across the entire cohort of patients, a discernible trend indicates a decrease in insulin requirements from mid-night to 6 a.m., with a more pronounced decline observed between 02:00 and 06:00. Beyond 6 a.m., there is a subse-quent increase in insulin requirements.
Similarly, in non-diabetic patients, insulin secretion tends to decrease during sleep. Specifically, in the first half of the night, a reduction in insulin production may be observed as the body adapts to the diminished need for insulin in the absence of food intake. Subsequently, towards the early morning hours, endogenous insulin production may gradu-ally increase in anticipation of the expected metabolic activ-ity upon waking.
AHCL systems, by mimicking the physiological produc-tion of insulin, allow for optimal management of nocturnal A retrospective study analyzed overnight, daytime, and total insulin requirement in 32 adults with T1DM showing that overnight insulin requirements were significantly more variable (CV 31%) than daytime amounts (CV 22%) [18].
In our study the observation of an interindividual coef-ficient of hourly insulin requirement greater than 50% in 14 nights of observation confirms the mechanism that leads to the superiority of the aHCL system compared to traditional MDI therapy and explains the frequent difficulty of people with type diabetes 1 to obtain good glycemic control with-out hypoglycemia during the nocturnal period.
The main limitations of our real-life study are the retro-spective design and the low number of participants. A ran-domized prospective study is needed to confirm our data.
In conclusion, all the three AHCL systems were equally able to achieve the recommended targets for CGM metrics overnight in adults with T1DM. The Tandem Control-IQ system obtained a higher tight TIR.
Acknowledgements The authors thank the participants whose data were included in this study.
Funding This work was partially funded by the National Plan for NRRP Complementary Investments (PNC, established with the de-cree-law 6 May 2021, n. 59, converted by law n. 101 of 2021) in the call for the funding of research initiatives for technologies and inno-vative trajectories in the health and care sectors (Directorial Decree n. 931 of 06-06-2022)-project n. PNC0000003-AdvaNced Technolo-gies for Human-centrEd Medicine (project acronym: ANTHEM). This work reflects only the authors’ views and opinions, neither the Minis-try for University and Research nor the European Commission can be considered responsible for them.
Data availability All data sets generated during and/or analyzed dur-ing the present study are available from the corresponding author on reasonable request.
Declarations
Conflict of interest No competing financial interests exist.
Informed consent Written informed consent to use the clinical and biochemical data was obtained from all participants.
Ethical standard statement All procedures performed in this study in-volving human participants were in accordance with the Ethical Stan-dard of the Institutional and National Research Committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standard. The study was approved by the Local Ethical Com-mittee.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.o rg/licenses/by/4.0/.



This article is excerpted from the 《Acta Diabetologica》 by Wound World.